Flow Neuroscience FL-100

Flow Neuroscience FL-100 aruschman Aug 13, 2026 Release DateAugust 13, 2026 Description

View CSAF

Summary

Successful exploitation of this vulnerability could allow an attacker within Bluetooth range to manipulate brain stimulation parameters and override safety limits.

The following versions of Flow Neuroscience FL-100 are affected:

  • Flow Neuroscience FL-100
  • Halo Neuroscience FL-100
CVSSVendorEquipmentVulnerabilities
v3 8.1Flow NeuroscienceFlow Neuroscience FL-100Use of Hard-coded Credentials

Background

  • Critical Infrastructure Sectors: Healthcare and Public Health
  • Countries/Areas Deployed: Worldwide
  • Company Headquarters Location: Sweden

Vulnerabilities

Expand All +

CVE-2026-18164

An undocumented hard-coded credential, shared by all device units, is authorized to bypass authentication. This allows an attacker within Bluetooth range to arbitrarilymanipulate brain stimulation parameters and state.

View CVE Details

Affected Products

Flow Neuroscience FL-100

Vendor:
Flow NeuroscienceProduct Version:
Flow Neuroscience Flow Neuroscience FL-100: <July_2026, Flow Neuroscience Halo Neuroscience FL-100: <July_2026Product Status:
known_affected

Remediations

Mitigation
Users are encouraged to install the latest firmware updates provided by Flow Neuroscience via the Flow app.

Relevant CWE:CWE-798 Use of Hard-coded Credentials

Metrics

CVSS VersionBase ScoreBase SeverityVector String
3.18.1HIGHCVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:H
4.07.2HIGHCVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:H/VA:H/SC:N/SI:N/SA:N

Acknowledgments

  • A.C. Buglione reported this vulnerability to CISA

Legal Notice and Terms of Use

This product is provided subject to this Notification (https://www.cisa.gov/notification) and this Privacy & Use policy (https://www.cisa.gov/privacy-policy).

Recommended Practices

CISA recommends users take defensive measures to minimize the risk of exploitation of this vulnerability.

Minimize network exposure for all control system devices and/or systems, ensuring they are not accessible from the internet.

Locate control system networks and remote devices behind firewalls and isolating them from business networks.

When remote access is required, use more secure methods, such as Virtual Private Networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as the connected devices.

CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.

CISA also provides a section for control systems security recommended practices on the ICS webpage on cisa.gov/ics. Several CISA products detailing cyber defense best practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.

CISA encourages organizations to implement recommended cybersecurity strategies for proactive defense of ICS assets.

Additional mitigation guidance and recommended practices are publicly available on the ICS webpage at cisa.gov/ics in the technical information paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.

Organizations observing suspected malicious activity should follow established internal procedures and report findings to CISA for tracking and correlation against other incidents.

CISA also recommends users take the following measures to protect themselves from social engineering attacks:

Do not click web links or open attachments in unsolicited email messages.

Refer to Recognizing and Avoiding Email Scams for more information on avoiding email scams.

Refer to Avoiding Social Engineering and Phishing Attacks for more information on social engineering attacks.

No known public exploitation specifically targeting this vulnerability has been reported to CISA at this time. This vulnerability is not exploitable remotely.

Revision History

  • Initial Release Date: 2026-08-13
DateRevisionSummary
2026-08-131Initial Publication

Legal Notice and Terms of Use

Advisory TypeICS Medical Advisory

TopicsIndustrial Control System Vulnerabilities, Industrial Control Systems

Alert CodeICSMA-26-225-01

SectorHealthcare and Public Health Sector

Show 'Key Takeaways' blockOff

On

添加评论
点赞收藏
点踩分享查看原文
评论
?
参与讨论